| Home > Publications database > Double-hit separation and dE/dx resolution of a time projection chamber with GEM readout |
| Typ | Amount | VAT | Currency | Share | Status | Cost centre |
| Hybrid-OA | 0.00 | 0.00 | EUR | (Publish and Read) | 289 | |
| Sum | 0.00 | 0.00 | EUR | |||
| Total | 0.00 |
| Journal Article | PUBDB-2022-03916 |
; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ;
2022
IOP Publishing
Bristol, UK
This record in other databases:
Please use a persistent id in citations: doi:10.1088/1748-0221/17/11/P11027 doi:10.3204/PUBDB-2022-03916
Report No.: DESY-22-089; arXiv:2205.12160
Abstract: A time projection chamber (TPC) with micropattern gaseous detector (MPGD) readout is investigated as main tracking device of the International Large Detector (ILD) concept at theplanned International Linear Collider (ILC). A prototype TPC equipped with a triple gas electron multiplier (GEM) readout has been built and operated in an electron test beam. The TPC wasplaced in a 1 T solenoidal field at the DESY II Test Beam Facility, which provides an electron beam up to 6 GeV/c. The performance of the readout modules, in particular the spatial point resolution, is determined and compared to earlier tests. New studies are presented with first results on the separation of close-by tracks and the capability of the system to measure the specific energy loss dE/dx. This is complemented by a simulation study on the optimization of the readout granularity to improve particle identification by dE/dx.
|
The record appears in these collections: |
Preprint
Double-hit separation and dE/dx resolution of a time projection chamber with GEM readout
(2022) [10.3204/PUBDB-2022-02594]
Files
Fulltext by arXiv.org
BibTeX |
EndNote:
XML,
Text |
RIS