| Home > Publications database > A Fiber detector radiation hardness test |
| Report/Journal Article | PUBDB-2017-10900 |
; ; ;
2000
North-Holland Publ. Co.
Amsterdam
This record in other databases:
Please use a persistent id in citations: doi:10.1016/S0168-9002(99)01437-0
Report No.: DESY-99-079; physics/9907019
Abstract: An intense 146 MeV /c pion beam was stopped inside a scintillating fiber detector made out of 12 planes with 16 pixels each, where every pixel consists of 8×8 densely packed scintillating fibers of 500 μm diameter. The detector was irradiated for 52 h to more than 10 kGy at its center. Before and directly after the irradiation the detector had been exposed to a particle beam to compare the corresponding light output. This study was continued during the following three months using cosmic rays. No damage was found taking into account the measurement errors of 5–10%. A 9 cm deep lucite degrader became irreversibly non-transparent in the irradiation region.
Keyword(s): scintillation counter: fibre ; radiation: damage ; pi: radiation ; HERA-B ; DESY HERA Stor
|
The record appears in these collections: |